|
(b)(4).Date sent: 9/25/2026.D4: batch #714d84.Investigation summary:
the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one psee60a device was returned with no apparent damage and with no reload present.In addition, the knife was noted to be partially advanced into the anvil, thus the device would not closed.The knife was returned to home position and the device was tested for functionality in the straight position with a test reload and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the remaining staples met the staple form release criteria.As part of our quality process, the manufacturing records of this batch number were reviewed, and the manufacturing standards were met prior to the release of this batch
as part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.No conclusion could be reached as to how the knife was partially advanced.A manufacturing record evaluation was performed for the finished device psee60a lot number a98m6x and no non-conformances were identified.A manufacturing record evaluation was performed for the finished device batch number 714d84, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|