Synthes GmbH Retention Pin; Screwdriver Back to Search Results
Catalog Number 03.045.008
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/08/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported during the procedure, the tip of the short quick anchor retention pin fractured.The fractured fragment remained lodged in the head of the locking screw for the nail.Another screw was used to complete the procedure successfully.It was also noted that a drill bit showed significant wear and loss of sharpness.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11: additional narrative: h3, h6: visual inspection of the photo found the device to be broken from the threaded tip.The broken fragment remains stuck on the screw head.No other anomalies were noted.The reported complaint was confirmed based on the evaluation results.The investigation traced the potential cause to a random component failure that may have been caused by exposure to unintended forces.It has been determined that no corrective or preventive actions are proposed.There is no indication that a design or manufacturing issue caused the reported condition.As part of depuy synthes quality processes, all devices are manufactured, inspected, and released in accordance with approved specifications.Additional monitoring for potential safety signals is conducted through complaint trend analysis as part of post-market safety surveillance activities.H3, h6: for complaints for which a non-manufacturing related probable cause has been identified no manufacturing record evaluation is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Retention Pin
Common Device Name
Screwdriver
Manufacturer (Section D)
Synthes GmbH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
Monument
1101 synthes avenue
monument CO 80132
Manufacturer Contact
elba bello
1302 wright lane east
west chester, PA 19380
+1(908)386-3534
MDR Report Key50898837
Report Number8030965-2026-09793
Device Sequence Number20333738
Product Code HXX
UDI-Device Identifier07612334167859
UDI-Public(01)07612334167859(10)427P980
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
Exempt
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.045.008
Device Lot Number427P980
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 09/08/2026
Initial Report FDA Received Date09/25/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/02/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-