| Catalog Number |
DYNJSC301 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/04/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that stopcocks are leaking.
|
| |
|
Additional Manufacturer Narrative
|
|
The account reported finding stopcocks are leaking.No sample or photo was received for investigation; finished good production records were reviewed and there were no line down situations, deviations, run-withouts, or non-conformance reports that might have had an impact on this issue, and the finished good pack build was reviewed.Because no sample or photo was received, the reported issue could not be evaluated on a returned sample; medline noted that the item is received from the vendor and is not altered during the kit assembly process.The component manufacturing facility identified that the leaking is caused by a mold deviation which leads to uneven wall thickness of the item.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|
| |
|
Search Alerts/Recalls
|
|
|