| Catalog Number |
20802010001 |
| Medical Device Problem Code |
Unintended System Motion (1430)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/15/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The subject device was evaluated by a field service engineer (fse).It was confirmed that rosa showed multiple collision errors and when moving the robotic arm to cut position it was shaking.Fse checked ft (force torque) sensor cable and saw the voltage would spike every time the cable was touched.To resolve the issue the ft sensor cable was replaced.The system was confirmed to be fully functional.The serial number of this device was reviewed for any deviations and/or anomalies in the servicing/maintenance process that may have contributed to this reported event.No deviations or anomalies were discovered therefore, it is not expected that the servicing/maintenance process contributed to the reported issue.The root cause cannot be determined with the information provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
It was reported that during surgery the system generated multiple collision errors and the arm shook when moved to the cut position.There was no patient involvement.The force torque sensor cable was found to be the cause of this unintended movement.
|
| |
|