| Model Number |
4473 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); High impedance (1291)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this right atrial (ra) lead exhibited high out of range pacing impedance measurements.The rep confirmed the impedances spiked out of range about three times over the past year and then returned to normal.Technical services (ts) recommended isometrics to test the lead integrity.There was noise present that was not oversensed.This lead has been recommended for replacement.This ra lead remains in service.No adverse patient effects were reported.
|
| |
|