The following has been reported: there is a crack on bottom-left of the front enclosure.A preliminary review of the device logs was not performed.The device was returned for evaluation.The front housing is damaged on the left side bottom corner and extending from the handle assembly to front face.During investigation it was found the cycle counts on the batteries were high (>115).While the batteries are not a source of failure at this time, they will be removed and replaced with new batteries.A broken screw mount was identified.Braided wire was present instead of insulated wire.The bag hook was missing.The lip seals had debris.During repair, the device experience a failing dsps calibration test (gain out of range).The dsps sensor was removed and replaced.The dsps calibration test was performed and the unit passed.After repair, the device passed all functional testing.An active infusion was not stopped.Reporting as a conservative measure for the downstream pressure sensor issue identified during investigation.No adverse effects were reported.
|