| Model Number |
72404257 |
| Medical Device Problem Codes |
Collapse (1099); Inflation Problem (1310); Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/16/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this inflatable penile prosthesis (ipp) stopped pumping up and stopped rebounding.A surgical procedure was performed during which the entire device was explanted and replaced with a malleable device.Upon explant the physician observed the right cylinder had a small pinhole.The physician believed the patient may have attempted augmenting themselves post-surgery.There were no patient complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Model number/catalog number: 720185-01.Serial number: n/a.Batch/lot number: 1100497844.Model/catalog description: reservoir flat iz 100 ml.Unique identifier (udi) #: (b)(4).
|
| |
|
Search Alerts/Recalls
|