MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; SNAPSECURE 3-WAY FOLEY DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; SNAPSECURE 3-WAY FOLEY DEVICE Back to Search Results
Catalog Number SNAPSECURE3W
Medical Device Problem Code Difficult to Open or Close (2921)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/06/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "foley securement device broken upon removal from package and did not click closed.".
 
Additional Manufacturer Narrative
The reported product problem/issue was broken upon removal from package, and the foley securement device did not click closed.Medline quality received 1 ea of the reported product in unpackaged, used condition; visual assessment revealed that one of the hooks was shorter than the other, and functional testing was conducted by closing the securement device.The reported issue of the snapsecure not closing was verified, and the device was not able to close due to the hooks not latching.Based on the provided sample and the description of the issue, the likely root cause is a manufacturing defect resulting in malformed hooks; based on the findings of the quality investigation, no change has been made to the regulatory reportability decision for the complaint.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
SNAPSECURE 3-WAY FOLEY DEVICE
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
raveena sampathi
three lakes drive
northfield, IL 60093--275
MDR Report Key50899152
Report Number1417592-2026-51119
Device Sequence Number16142061
Product Code EYJ
UDI-Device Identifier20193489102717
UDI-Public20193489102717
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberSNAPSECURE3W
Device Lot Number67024044502
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer01/06/2025
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/06/2024
Initial Report FDA Received Date09/25/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-