Medline Industries, LP MEDLINE; IV START KIT W/CHLORAPREP APP Back to Search Results
Catalog Number DYND74260
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/21/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The customer reported the same problem with the tegaderm in the iv start kit: the film is difficult to take off with the tabs, and the film is staying on the backing and not able to be peeled off correctly.A photo and a sample were received; sample evaluation found 3 dressings without the rest of the kits, with 2 dressings returned as just the liners and the 3rd dressing stuck to the liner with the paper frame separated from the dressing so that it could no longer be used.Finished good production records and finished good pack build were reviewed, and there were no run without approvals, deviations or non-conformance reports and no issues with the build that likely contributed to the reported issue.Medline confirmed the vendor and the cause of this issue; the issue derives from the raw materials process when the force between frame liner and the film is too weak, and medline is working with the component manufacturing facility for further evaluation and corrective actions.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Event or Problem Description
It was reported that "the film is difficult to take off with the tabs.The film is staying on the backing and not able to be peeled off correctly.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
IV START KIT W/CHLORAPREP APP
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
joseph weiske
three lakes drive
northfield, IL 60093--275
MDR Report Key50899162
Report Number1417592-2026-51122
Device Sequence Number18830959
Product Code LRO
UDI-Device Identifier40080196992981
UDI-Public40080196992981
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberDYND74260
Device Lot Number24BBL165
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer03/29/2024
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/21/2024
Initial Report FDA Received Date09/25/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1