Model Number U225
Medical Device Problem Codes Failure to Capture (1081); Pacing Problem (1439)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/06/2026
Type of Reportable Event Malfunction
Event or Problem Description
T was reported that this cardiac resynchronization therapy pacemaker (crt-p) and the left ventricular (lv) lead suspended a left ventricular automatic threshold (lvat) test.Technical services (ts) reviewed device data and confirmed the lvat was suspended due to fusion beats.Also, the presenting electrogram (egm) exhibited loss of capture on the lv lead due to the lv and right ventricular (rv) beats falling on top of each other.Ts recommended further evaluation in clinic.This crt-p remains in service.No adverse patient effects were reported.
 
Additional Manufacturer Narrative
This product investigation is still in progress.This report will be updated when product investigation is complete.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISIONIST CRT-P
Common Device Name
Pulse Generator, Pacemaker, Implantable, With Cardiac Resynchronization (Crt-P)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
Boston Scientific Corporation
cashel road
clonmel E91 T 862
EI   E91 T862
Manufacturer Contact
mark mischke
4100 hamline avenue north
saint paul, MN 55112
+1(612)875-7234
MDR Report Key50899270
Report Number2124215-2026-51231
Device Sequence Number10761514
Product Code NKE
UDI-Device Identifier00802526559426
UDI-Public(01)00802526559426(17)280522
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberU225
Device Catalogue NumberU225
Device Lot Number736751
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/11/2026
Initial Report FDA Received Date09/25/2026
Date Device Manufactured06/02/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age88 YR
Patient SexMale
-
-