| Model Number |
U225 |
| Medical Device Problem Codes |
Failure to Capture (1081); Pacing Problem (1439)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/06/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
T was reported that this cardiac resynchronization therapy pacemaker (crt-p) and the left ventricular (lv) lead suspended a left ventricular automatic threshold (lvat) test.Technical services (ts) reviewed device data and confirmed the lvat was suspended due to fusion beats.Also, the presenting electrogram (egm) exhibited loss of capture on the lv lead due to the lv and right ventricular (rv) beats falling on top of each other.Ts recommended further evaluation in clinic.This crt-p remains in service.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This product investigation is still in progress.This report will be updated when product investigation is complete.
|
| |
|
Search Alerts/Recalls
|