|
The customer reported that the point seems smaller and flimsy and it does not work for removing regular skin staples as it is too small.No sample or photo was received for the investigation; finished good production records were reviewed, and there were no run without approvals, deviations or non-conformance reports that might have had an impact on this issue, and the finished good pack build was reviewed with 1 ea of staple remover present in the kit.At this time, medline is unable to confirm the reported defect, as no physical sample or detailed photographic evidence was provided to support the investigation.The cause was unconfirmed, sample required to determine cause, and the division will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|