• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Abbott Medical HF Sensor Delivery System; System, Hemodynamic, Implantable

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Abbott Medical HF Sensor Delivery System; System, Hemodynamic, Implantable Back to Search Results
Model Number CM2000
Medical Device Problem Code Incorrect Measurement (1383)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/22/2026
Type of Reportable Event Serious Injury
Event or Problem Description
A right heart catheterization was performed to confirm the validity of the pressure sensor readings.The sensor was not recalibrated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF Sensor Delivery System
Common Device Name
System, Hemodynamic, Implantable
Manufacturer (Section D)
Abbott Medical
387 nerem st nw suite 500
atlanta GA 30313
Manufacturer (Section G)
Abbott Medical
387 nerem st nw suite 500
atlanta GA 30313
Manufacturer Contact
fredric gold
168 middlesex turnpike
burlington, MA 01803
+1(651)756-5400
MDR Report Key50899300
Report Number3004936110-2026-01665
Device Sequence Number10705751
Product Code MOM
UDI-Device Identifier05414734509237
UDI-Public(01)05414734509237(10)9151351(17)250803
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P100045
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/03/2025
Device Model NumberCM2000
Device Catalogue NumberCM2000
Device Lot Number9151351
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/22/2026
Initial Report FDA Received Date09/25/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/04/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age78 YR
Patient SexFemale
Patient Weight182 LB
-
-