It was reported that upon completion of the product analysis, this device was confirmed to have been operating in safety mode.The device was confirmed to be able to perform a memory dump, and it never triggered the elective replacement indicator.No additional adverse patient effects were reported.This device is no longer in service.
|
|
Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.Analysis of the device confirmed it was operating in safety mode and that critical therapy remained available.Engineers determined that this device was demonstrating behavior consistent with a high battery impedance, which put this device at risk of experiencing transient voltage decreases (and associated resets) during telemetry or other normal, higher-power operations.When battery voltage drops below a minimum threshold, a system reset is performed.If three system resets occur within a 48-hour period, the device is designed to enter safety mode operation to maintain back-up pacing with pre-defined settings.Boston scientific has issued a field safety notice to notify physicians of the potential for accolade family pacemaker and cardiac resynchronization therapy pacemaker (crt-p) devices to exhibit this behavior.A review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.
|