The device history record (dhr) for the reported lot was reviewed and no issues were identified.The product was manufactured, tested, and released in accordance with validated procedures and established acceptance criteria.Visual inspection of the returned device revealed damage incurred upon explantation.Due to this damage, the device was unable to undergo functional testing, including the steady state pressure test that mimics physiological ocular flow conditions.The valve was successfully primed without issue.Based on the evaluation performed, the reported issue could not be replicated or confirmed, and a manufacturing-related cause could not be established.
|