| Catalog Number |
309657 |
| Medical Device Problem Code |
Contamination /Decontamination Problem (2895)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/14/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that bd syringe 3ml ll 200 s/c contained foreign matter.Verbatim:
complaint via email.On 14jul2026, during manufacturing a small, black free floating particulate was identified in the product.The particulate was sent out to a third party for identification which determined that it was polypropylene with silicone (see attached).After review of all the components used in the manufacturing process up to the point where the particulate was found, determined that several different syringes were used in the process (see below) which are comprised of polypropylene and silicone.X would like to request an investigation into these three lots of syringes to determine if there were any issues during manufacturing, along with controls in place to prevent particulate contamination of the syringes specifically polypropylene and/or silicone.Please let me know if there are any questions
regards:
nvs sap material: 5004641.Nvs sap batch: mp0018.Vendor: 50085232 / becton dickinson.Vendor material #: 309657.Vendor batch: 4285707.Expiry: 30sep2029.Po # (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
|
| |
|