MAUDE Adverse Event Report: C.R. Bard Inc. (Covington) -1018233 Bardex® Foley Catheter Silicone Coated

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. Bard Inc. (Covington) -1018233 Bardex® Foley Catheter Silicone Coated Back to Search Results
Catalog Number 0167V22S
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Hematuria (2558)
Date of Event 09/17/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that following transurethral resection of a bladder tumor, a balloon catheter was left in place to facilitate continuous bladder irrigation.During treatment, it was observed that the continuous bladder irrigation drainage was obstructed, and the irrigation fluid could not drain out smoothly.The medical team immediately conducted a bedside examination and confirmed that the catheter had dislodged and the balloon at its tip had ruptured.The faulty catheter was promptly removed, the patient¿s urethra and bladder were assessed, a new catheter was inserted, and continuous bladder irrigation was resumed.The patient was closely monitored for urine color, bleeding volume, urethral injury, and worsening hematuria.Subsequently, the patient¿s condition remained relatively stable, with no serious complications such as massive bleeding or urethral laceration.However, this incident prolonged the duration of the surgical procedure and increased the duration of anesthesia.Furthermore, fragments from the ruptured balloon posed a risk of remaining in the bladder as foreign bodies, potentially leading to secondary infection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Bardex® Foley Catheter Silicone Coated
Common Device Name
Foley Catheter
Manufacturer (Section D)
C.R. Bard Inc. (Covington) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. Bard Inc. (Covington) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington, GA 30014
+1(770)784-6100
MDR Report Key50899375
Report Number1018233-2026-06106
Device Sequence Number13439745
Product Code KOD
UDI-Device Identifier00801741019234
UDI-Public(01)00801741019234(17)280828(10)MYHX1021
Combination Product (Y/N)N
Initial Reporter StateNA
Initial Reporter CountryCH
PMA/510(K) Number
K922431
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number0167V22S
Device Lot NumberMYHX1021
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/21/2026
Initial Report FDA Received Date09/25/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-