| Model Number |
480275-12 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/02/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted procedure, a screw on a harmonic ace instrument was observed to have loosened or broken and a fragment fell inside the patient.The fragment was retrieved during the same surgery.The customer stated that the broken fragment was immediately retrieved using a long forceps instrument along with another robotic arm.To confirm complete retrieval, the surgical site was thoroughly irrigated and carefully inspected using the da vinci robotic endoscope, ensuring no fragments remained in the area.A postoperative ultrasound examination was conducted following the visual confirmation that no residual fragment was present.
|
| |
|
Additional Manufacturer Narrative
|
|
The harmonic ace instrument has not been received for failure analysis evaluation.
|
| |
|