| Model Number |
106525US |
| Medical Device Problem Code |
Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/07/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the patient had a driveline communication fault alarm and presented to the clinic for evaluation.The modular cable was exchanged, which the patient tolerated without issue.Log files were obtained before the exchange.The log file captured driveline communication fault alarms beginning at 5:55 am the morning of (b)(6) 2026.The system continued to operate at the set speed and was supported by the controller¿s backup battery until external power was restored.There were no other unusual events recorded in the log file event history.The mechanical circulatory support (mcs) equipment was operating as intended.The alarm resolved after the modular cable was exchanged.
|
| |
|
Additional Manufacturer Narrative
|
|
No further information was provided.A supplemental report will be submitted once the manufacturer¿s investigation is complete.
|
| |
|