| Model Number |
CDS0801-XT |
| Medical Device Problem Code |
Incomplete Coaptation (2507)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2026 a patient presented with grade 4+ functional mitral regurgitation (mr), rotated heart and restricted anterior leaflet for a mitraclip procedure.During the procedure, mitraclip was implanted.After deployment, it was determined that there was a single leaflet device attachment (slda) and the clip was no longer attached to the anterior leaflet.Additional clip was implanted to stabilize the slda.Imaging was difficult.The final mr was grade 1-2.There were no adverse patient effects or clinically significant delays reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history did not indicate a lot-specific quality issue.Based on available information, the reported slda was due to patient anatomy (rotated heart).The cause of the reported difficult imaging was unable to be determined.The reported unexpected medical intervention was the result of case-specific circumstances.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
| |
|