| Model Number |
TCDS0801-XTW |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2026 a patient presented with grade 5 functional tricuspid regurgitation (tr) for a triclip procedure.During the procedure, first establish final arm angle (efaa) was passed, lockline was removed.During second first establish final arm angle (efaa), clip failed.Clip was removed from the patient and was replaced with another clip.All steps were performed as per ifu.The final tr was grade 2.There were no adverse patient effects or clinically significant delays reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
|
| |
|