Medline Industries, LP MEDLINE; THERMOMETER, TALKING, EAR/FOREHEAD, HOME Back to Search Results
Catalog Number MDSTH1002
Medical Device Problem Codes Computer Software Problem (1112); Failure to Shut Off (2939)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/27/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "put the batteries in it and it hasnt turned off since, the screen does not go black, the last tempature reading and time stays on 24/7 unless batteries are taken out.".
 
Additional Manufacturer Narrative
The customer reported the thermometer "hasnt turned off since," "the screen does not go black," and "the last tempature reading and time stays on 24/7 unless batteries are taken out." medline received the thermometer and, upon functional testing, found it did respond to all function that were input by using the buttons it provides; however, it did stay on at all time, and as soon as the batteries went in to the device it would stay on till batteries were removed.The complaint was confirmed.The updated cause code is defect confirmed: unknown responsible party, cannot determine cause w/received sample; no further action was taken because objectionable trends were not seen for this issue, and division has been informed and will continue to monitor trends.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
THERMOMETER, TALKING, EAR/FOREHEAD, HOME
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
shirisha mothe
three lakes drive
northfield, IL 60093--275
MDR Report Key50899650
Report Number1417592-2026-51185
Device Sequence Number5829100
Product Code FLL
UDI-Device Identifier00080196406404
UDI-Public00080196406404
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K100488
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberMDSTH1002
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/27/2023
Initial Report FDA Received Date09/25/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-