The customer reported the thermometer "hasnt turned off since," "the screen does not go black," and "the last tempature reading and time stays on 24/7 unless batteries are taken out." medline received the thermometer and, upon functional testing, found it did respond to all function that were input by using the buttons it provides; however, it did stay on at all time, and as soon as the batteries went in to the device it would stay on till batteries were removed.The complaint was confirmed.The updated cause code is defect confirmed: unknown responsible party, cannot determine cause w/received sample; no further action was taken because objectionable trends were not seen for this issue, and division has been informed and will continue to monitor trends.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|