| Model Number |
BF-P190 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer returned the bronchovideoscope to the service center for a separate issue.During the device analysis, the investigator indicated the objective lens is missing glue, and the light guide lens is chipped.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection and the reported failure was identified.Based on the results of the investigation, a root cause could not be identified.Adhesive issue: reasonably known information suggests probable causes such as material problems like: degredation.Mechanical problems like: device migration; stress; wear and tear; leakage; lubrication.These causes can occur between components such as mesh; tube; cover; label; cannula; coating material; power cord; electrical cable; tip; adhesive; lenses; electrode; housing; stain relief; handpiece; access port; camera; and connector/coupler; knob; ring; prong; tube; joint; switch/relay; connector pin; plate; and lenses without any design or manufacturing issue.That could lead to apperance issues.Lens issue: reasonably known information suggests probable causes such as material problems like: degredation.Mechanical problems like: device migration; stress; wear and tear; leakage; lubrication.These causes can occur between components such as mesh; tube; cover; label; cannula; coating material; power cord; electrical cable; tip; adhesive; lenses; electrode; housing; stain relief; handpiece; access port; camera; and connector/coupler; knob; ring; prong; tube; joint; switch/relay; connector pin; plate; and lenses without any design or manufacturing issue.That could lead to apperance issues.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|