| Model Number |
107754 |
| Medical Device Problem Code |
Complete Loss of Power (4015)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/29/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the patient had many no external power alarms.The patient stated that they pull on the mobile power unit (mpu) when going to the restroom and wanted to verify how often the event occurred.The log files were submitted for review.No external power events were recorded on (b)(6) 2026 at 1:01:17 and on (b)(6) 2026 at 3:11:32 that was likely caused by power to the mpu being briefly interrupted.There was another no external power event recorded on (b)(6) 2026 at 20:23:13 that appeared to have been accidental as the patient was attempting to switch from the mpu to battery power.The emergency backup battery (ebb) was used to support the system during the events.The ebb use count was 7.
|
| |
|
Additional Manufacturer Narrative
|
|
No further information was provided.A supplemental report will be submitted once the manufacturer¿s investigation is complete.
|
| |
|