| Model Number |
2120 |
| Medical Device Problem Code |
Device Sensing Problem (2917)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/15/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was found during the investigation of a returned pump that the customer complaint regarding the device displaying the error message, "cassette not attached properly.Reattach cassette," despite the cassette being properly attached, was confirmed through investigation testing.This error code triggers a high-priority alarm, and this issue could potentially interrupt therapy.There was no patient involvement and no harm was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspected device was returned for evaluation.The review of event log performed and found nothing related to complaint.A review of the available service history identified no previous related repairs within the last 12 months.The visual inspection and functional testing were performed, and defective downstream occlusion (dso) sensor was found and replaced.The customer complaint was confirmed.Additionally, the investigation identified a damaged uso seal, battery compartment, and defective main board, all of which were replaced.The downstream occlusion (dso) and upstream occlusion (uso) sensors were calibrated.The device passed all functional testing after repair.The software was updated, and the device was reset to its standard configuration.
|
| |
|