• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: St. Jude Medical - Neuromodulation (Puerto Rico, LLC) Proclaim Plus Implantable Pulse Generator; Stimulator, Spinal-Cord, Totally Implanted For Pain Relief

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

St. Jude Medical - Neuromodulation (Puerto Rico, LLC) Proclaim Plus Implantable Pulse Generator; Stimulator, Spinal-Cord, Totally Implanted For Pain Relief Back to Search Results
Model Number 3670
Medical Device Problem Code Wireless Communication Problem (3283)
Health Effect - Clinical Code Inadequate Pain Relief (2388)
Date of Event 08/28/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that following an unrelated surgical procedure wherein the device was not placed into surgery mode, the ipg became unable to communicate with external devices.Troubleshooting has been unable to resolve the issue.As a result, surgical intervention may be undertaken to address the issue.
 
Additional Manufacturer Narrative
Date of event is estimated.The device is included in the neuromodulation implantable pulse generator (ipg) proclaim¿ scs family, proclaim¿ drg therapy and infinity¿ dbs systems surgery mode advisory notice issued by abbott on 02 april 2026.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Proclaim Plus Implantable Pulse Generator
Common Device Name
Stimulator, Spinal-Cord, Totally Implanted For Pain Relief
Manufacturer (Section D)
St. Jude Medical - Neuromodulation (Puerto Rico, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer (Section G)
St. Jude Medical - Neuromodulation (Puerto Rico, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer Contact
sara speyerer
6901 preston road
plano, TX 75024
+1(972)309-8000
MDR Report Key50899863
Report Number3006705815-2026-05815
Device Sequence Number14910188
Product Code LGW
UDI-Device Identifier05415067046383
UDI-Public(01)05415067046383
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
P010032
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Other,Recall,Notification
Type of Report Initial,Followup
Report Date (Section B) 09/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/25/2024
Device Model Number3670
Device Catalogue Number3670
Device Lot NumberA000132995
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/28/2026
Supplement Date Received by Manufacturer09/30/2026
Initial Report FDA Received Date09/25/2026
Supplement Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/26/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
1627487-04/02/26-001-C
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SCS Anchor (x2).; SCS Lead (x2).
Outcome Attributed to Adverse Event Other;
Patient SexFemale
Patient Weight176 LB
-
-