MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. StealthStation¿ S8 system; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. StealthStation¿ S8 system; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 9735665
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/06/2026
Type of Reportable Event Malfunction
Event or Problem Description
Medtronic received information regarding a navigation system being used intra-operatively during a cervical spine procedure.It was reported that the site felt they were 5mm off (inferior) after the spin.They swapped the frame clamp it was still off by 5mm (inferior) even after re-spinning with a new clamp. as they worked towards the frame they found that accuracy improved, but was never perfect.There was a less than one hour surgical delay and no impact on patient outcome.
 
Additional Manufacturer Narrative
Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id: 9735737, serial/lot #: 2.1.0, ubd: , udi#: h3, h6: a medtronic representative went to the site to test the equipment.Testing revealed that there were no failures found.The s ystem performed as intended.Codes b01, c19, and d14 are applicable to this analysis.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
StealthStation¿ S8 system
Common Device Name
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC.
200 medtronic drive
lafayette CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC.
200 medtronic drive
lafayette CO 80027
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
+1(763)526-5677
MDR Report Key50899891
Report Number1723170-2026-01375
Device Sequence Number20333768
Product Code HAW
UDI-Device Identifier00643169722187
UDI-Public(01)00643169722187(11)190213
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K162309
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9735665
Device Catalogue Number9735665
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/08/2026
Initial Report FDA Received Date09/25/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/13/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
"See H11...."
Patient Age67 YR
Patient SexMale
Patient Weight237 LB
-
-