MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION MOMENTUM? EL ICD VR; Implantable Cardioverter Defibrillator

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION MOMENTUM? EL ICD VR; Implantable Cardioverter Defibrillator Back to Search Results
Model Number D120
Medical Device Problem Codes Premature Discharge of Battery (1057); Low impedance (2285)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/11/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that this implantable cardioverter defibrillator (icd) system recorded a code 1004 indicative of a short circuit condition during shock delivery.Technical services (ts) reviewed the data and noted there was a ventricular fibrillation (vf) episode.A shock was delivered and converted the rhythm; however, it resulted in a short.Ts analysis also concluded the device did not record a code 1004, but the shock resulted in a low out of range shock impedance measurement.It was also noted this icd system recorded a code 1003, which states that the voltage is too low for projected remaining capacity.The code 1003 was recorded due to the low shocking lead impedance.Ts recommended replacing the entire icd system.It was noted the patient would see the physician to discuss possible system extraction.Additionally, the patient is currently wearing a life vest until future plans are solidified and that tachy therapy was turned off.No adverse patient effects were reported.This icd remains in service.
 
Additional Manufacturer Narrative
This product investigation is still in progress.This report will be updated when product investigation is complete.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOMENTUM? EL ICD VR
Common Device Name
Implantable Cardioverter Defibrillator
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
Boston Scientific Corporation
cashel road
clonmel E91 T 862
EI   E91 T862
Manufacturer Contact
mark mischke
4100 hamline avenue north
saint paul, MN 55112
+1(612)875-7234
MDR Report Key50899894
Report Number2124215-2026-51253
Device Sequence Number6001230
Product Code LWS
UDI-Device Identifier00802526587702
UDI-Public(01)00802526587702(17)260130
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
P960040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date01/30/2026
Device Model NumberD120
Device Catalogue NumberD120
Device Lot Number220330
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/11/2026
Initial Report FDA Received Date09/25/2026
Date Device Manufactured02/11/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age78 YR
Patient SexMale
-
-