Model Number CM1100
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/08/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Fa-q323-hf-4 cardiomems pes right angle power extension cable failure notice issued by abbott on 04 oct 2023.¿ one cardiomems patient electronic system was received for evaluation.Visual inspection verified the power extension cable was frayed and wires were exposed.The backplate of the device was removed, and the returned power supply was connected to the unit.The unit was powered on with no issues observed.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The review determined that no non-conformances were identified that are related to the reported event.
 
Event or Problem Description
Exposed and frayed wires of the power extension cable were discovered during evaluation of the device.The patient electronic unit was replaced and there were no adverse consequences reported by the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOMEMS PATIENT ELECTRONIC SYSTEM
Common Device Name
System, Hemodynamic, Implantable
Manufacturer (Section D)
Abbott Medical
387 nerem st nw suite 500
atlanta GA 30313
Manufacturer (Section G)
Abbott Medical
387 nerem st nw suite 500
atlanta GA 30313
Manufacturer Contact
fredric gold
168 middlesex turnpike
burlington, MA 01803
+1(651)756-5400
MDR Report Key50899921
Report Number3004936110-2026-01666
Device Sequence Number17618203
Product Code MOM
UDI-Device Identifier05414734509800
UDI-Public(01)05414734509800(10)7600059
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
P100045
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Consumer,Health Professional,Company Representative
Initial Reporter Occupation Nurse
Remedial Action Notification
Type of Report Initial
Report Date (Section B) 09/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberCM1100
Device Catalogue NumberCM1100
Device Lot Number7600059
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/15/2026
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/23/2026
Initial Report FDA Received Date09/25/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/27/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
FA-Q323-HF-4
Patient Sequence Number1
Patient SexUnknown
-
-