|
The account reported ra separated from manifold body.No sample was received for evaluation; a review of the product drawing and device history record for the component were performed and determined to meet product specifications, with no non-conformances identified that would have contributed to the reported defect.Without a photo or sample, the reported issue and root cause could not be confirmed at this time.Additional information was provided and it revealed this 200s manifold was being used outside the intended design specs, with a power injector; the manifold is not intended for use with power injection.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|