| Model Number |
WB91051W |
| Medical Device Problem Code |
Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported the generator had an e433 error during an unspecfied procedure.During the initial evaluation of the device it was observed, the generator board had failed.No patient harm was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection and the reported failures were¿confirmed.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunctions: an electrical related component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|