| Model Number |
e 411 |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/29/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
There was an allegation of questionable results for multiple assays from the cobas e411 rack analyzer.The samples were repeated on a cobas e 402 analyzer.For the anti-hav igm assay, the elecsys hbsag ii assay, and the elecsys hiv combi pt assay, the results were allegedly false positives.No specific data was provided.The questionable results were not reported outside of the laboratory.
|
| |
|
Additional Manufacturer Narrative
|
|
The anti-hav igm reagent lot number was 914965.The hbsag assay reagent lot number was 908905.The hiv combi pt reagent lot number was 922989.The expiration dates were not provided.A local biomedical technician inspected the machine and determined that the measuring cell needed to be replaced, and a new blank cell and a cell check were required.The investigation is ongoing.
|
| |
|
Search Alerts/Recalls
|