| Model Number |
9735665 |
| Medical Device Problem Code |
Positioning Problem (3009)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Medtronic received information regarding a navigation system being used during a cranial biopsy procedure.It was reported that automated trajectory guidance unit was unable to lock into target.The unit only showed the dashed circle indicating there was no reachability information available.A manufacturer representative (rep) confirmed the ip address and communication between navigation system and targeting unit, the tool verified.They were in the navigation task on the system and the surgical plan was available.The unit was in home position after gross positioning.There was a delay of less than 1 hour and no impact on patient outcome reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Continuation of d10: product id 9735670 (serial: (b)(6)); product type: ; implant date n/a; explant date n/a.H3: a medtronic representative went to the site to test the equipment.Testing revealed that no faults or failures were found.The navigation system then passed the system checkout and was found to be fully functional.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|