• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; BLUETOOTH BP MONITOR,UNIVERSAL CUFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; BLUETOOTH BP MONITOR,UNIVERSAL CUFF Back to Search Results
Catalog Number MDS8001B
Medical Device Problem Code High Readings (2459)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/10/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "item 7328 medline blood pressure monitor with bluetooth is not working properly.Member stated the readings are 25 points off.".
 
Additional Manufacturer Narrative
The customer reported receiving a medline blood pressure monitors with bluetooth that "readings are off"; the member stated the readings are "25 points off," went to her doctor, "they took her bp and it was normal," and "then took bp reading with her monitor and it was off by 25 points." at this time, a sample(s) was not received for investigation, and investigation included review of recent complaint trends/scar activity for the last three trend reports.The complaint was not confirmed, and no trends were identified for the specified material number and its corresponding defect type.No further action was taken because without the sample or picture, it is difficult to truly identify the root cause of the issue; the division will continue to monitor trends.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
BLUETOOTH BP MONITOR,UNIVERSAL CUFF
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
donovan frazier
three lakes drive
northfield, IL 60093--275
MDR Report Key50900146
Report Number1417592-2026-51235
Device Sequence Number12219791
Product Code DXN
UDI-Device Identifier00197344077079
UDI-Public00197344077079
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K210770
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberMDS8001B
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/10/2025
Initial Report FDA Received Date09/25/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-