The customer reported receiving a medline evencare g2 glucose meter that was giving low readings during use, and no other information was available other than "not reading correctly for the low range of the test solution." medline received the meter in used condition; the sample came with no batteries, the customer did not send in the control strips or solutions they used with the sample, and medline inserted 2 aaa control batteries and tested the sample according to the instructions for use using evencare g2 glucose control solutions and evencare g2 blood glucose test strips.All test results were within control solution range; low control solution readings were 44 mg/dl and 187 mg/dl for high control solution, and during testing medline experienced no issues.The complaint was not confirmed, and medline was unable to confirm the complaint issue or that the product failed to meet specifications.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|