B3: date of event is unknown.The date received by manufacturer has been used for this field.E.1.Initial reporter facility name: (b)(6) hospital
d.4.Medical device expiration date: unknown
h.4.Device manufacture date: unknown
the lot number was not reported; however, two potential lot numbers were provided.The information for that number is as follows:
d4.Medical device lot#: 6135940
d4.Medical device expiration date: 28-feb-2027
h4.Device manufacture date: 15-may-2026
d4.Medical device lot#: 6135939
d4.Medical device expiration date: 28-feb-2027
h4.Device manufacture date: 15-may-2026
investigation summary:
bd received 2 photos for investigation related to lot numbers 6135939, 6135940 and unknown.Upon evaluation, the photos did not show the customer¿s reported failure mode of overfill.A total of 100 retained samples from both lots were visually inspected, with no issues identified.Functional tests were performed on 10 retained samples from the production lots, and all tubes were found to be within specification limits.The device history records were reviewed with no issues being identified.There were no related quality notifications.All process and final inspections comply with specification requirements.Additionally, an unknown lot number was provided; therefore, a review of the device history record for the unknown lot number could not be conducted.This complaint has not been confirmed for the lots 6135939, 6135940, and unknown, for the indicated failure mode: draw volume.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for the identification of emerging trends.
|