| Model Number |
IntelliVue Patient Monitor MX800 |
| Medical Device Problem Code |
Incorrect Measurement (1383)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Philips received a complaint on the intellivue mx800 patient monitor indicating the bis module is faulty and displaying incorrect readings.The device was not in use on patient at time of event, there was no adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(6)
a philips remote service engineer (rse) spoke to the biomedical engineer (bme) who evaluated the device and replicated the issue by testing the bis module with a simulator.The bis module displayed incorrect readings, confirming that the bis engine was faulty.A nemo (non engineering material only) service was agreed upon.The customer ordered a replacement bis module to resolve the issue.It has been concluded that no further action is required at this time.If additional information is received the complaint file will be reopened.
|
| |
|