| Model Number |
ZM-531PA |
| Medical Device Problem Codes |
Unable to Obtain Readings (1516); Incorrect, Inadequate or Imprecise Result or Readings (1535); Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/03/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The biomedical engineer (biomed) reported that the zm transmitter was experiencing intermittent signal loss at the central nurse's station (cns), losing the signal for a few seconds to minutes at a time.No patient harm was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The biomedical engineer (biomed) reported that the zm transmitter was experiencing intermittent signal loss at the central nurse's station (cns), losing the signal for a few seconds to minutes at a time.Technical support (ts) asked that he check the rssi values for this zm device, but he stated that they were using it in hiq-view.Ts had him confirm that there was an antenna outside the room.The biomed removed the patient from the device and tested it on a simulator and the issue followed the zm device.No patient harm was reported.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.
|
| |
|