| Model Number |
480299-08 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/02/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted procedure, the tip of the suction irrigator instrument broke off, and a fragment fell into the patient's anatomy.The fragment was successfully retrieved during the same procedure.A backup instrument was used to complete the procedure, and no patient injury was reported.No additional procedure was required.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical inc.(isi), has not received the suction irrigator instrument to perform failure analysis.
|
| |
|