| Model Number |
381121-36 |
| Medical Device Problem Code |
Visual Prompts will not Clear (2281)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted surgical procedure that non recoverable fault 86 occurred.The customer hard power cycled the system and performed an emergency power off, but the error reoccurred.The customer converted the procedure to laparoscopic.There were no reports of patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation was completed to determine the cause of this reported event.An intuitive surgical, inc.(isi) field service engineer (fse) was dispatched to the customer site to further investigate the reported event.The fse noted that he replaced the intuitive surgical core power distribution (ipd) board with a "temporary" ipd board to correct the reported non recoverable fault.The system was tested and verified as ready for use.Isi did not receive a da vinci product involved with this complaint to perform failure analysis.The complaint was confirmed based on the field evaluation.Blank mdr fields:
the missing patient information in sections a and b was either unknown, unavailable, not provided, or not applicable.The expiration date for section d4 is not applicable.Field d6 is blank because the product is not implantable.Information for the blank fields in section e1 is not available.Fields g5 and g7 are not applicable.Field h10 is blank as there are no related report numbers.
|
| |
|