| Model Number |
AFR-00008 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/28/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during cardiac ablation procedure, issue with return 1 port was observed.It was noted that ¿return 1 contact-quality-invalid error¿ message was displayed.Grounding patches were replaced without resolve.Adapter and gen link cable were replaced without resolution.Additionally, the system was rebooted multiple times, and the cable was reseated which did not resolve the i ssue.The procedure was aborted and was not continued with another system.Service was requested.Technical services initially requested replacement of the return electrode adapter.Pulsed field generator (pfg) issue was verified during servicing, and it was determined that the issue was related to pulsed field generator (pfg) only.Patient grounding block sheathing with wires.High impedance was confirmed on port 1 of the pulsed field generator (pfg).Pfg assembly was replaced.The console was serviced as appropriate.No further patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|