| Model Number |
KR2110X15B |
| Medical Device Problem Codes |
Circuit Failure (1089); Battery Problem (2885); No Tactile Prompts/Feedback (4024)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The reported failure of internal battery failure and ventilator inoperative alarms is accommodated in the product risk management file as being linked to hazard with description loss of function/loss of primary function.Complaints for this failure are reviewed via the post market complaint trending and escalation processes to assess for the observed probability levels and compare them with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.Dhr review was performed for the complaint device (b)(6).Review confirms that the device met the final acceptance criteria and was released for final distribution.
|
| |
|
Event or Problem Description
|
|
A trilogy evo was returned to the manufacturer's service center for evaluation.There was no harm or injury reported.The device was not in patient use.During the evaluation a service required alarm condition indicating the internal battery failed.The device's internal battery was replaced to address the issue.The device alarmed for ventilator inoperative alarms indicating communication failures.The system board was replaced to address the issues.The device alarmed indicating a touchscreen failure.The device's display board was replaced to address the issue.The device logs were not available; therefore, the date of the most recent error codes occurrence could not be determined.At this time, the reported error code event date is based solely on the customer's information received and has not been confirmed through device evaluation.Additionally, not related to the reportable issue, the device alarmed indicating a detachable battery failure.The device's detachable battery was replaced to address the issue.This would not affect therapy delivery.The blower assembly was found to have corrosion and was replaced.There was no allegation of blower failure.The device failed an ambient light sensor verification test step.The front panel was replaced to address the issue.The device was tested and passed all final testing.
|
| |
|
Search Alerts/Recalls
|