| Model Number |
CM2000 |
| Medical Device Problem Code |
Incorrect Measurement (1383)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/10/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A right heart catheterization was performed to confirm the validity of the pressure sensor readings.The sensor was not recalibrated.
|
| |
|
Additional Manufacturer Narrative
|
|
No device analysis results are available because the device remains implanted.A review of the device history record was performed and ruled out the possibility of a manufacturing related issue causing or contributing to the reported complaint issue.Per the instructions for use, right heart catheterization is required for system baseline (mean pressure) calibration and may be needed to recalibrate the baseline (mean pressure) in order to continue to use the system.
|
| |
|