DePuy Ireland - 9616671 VELYS; Instrument, Surgical, Orthopedic, Ac-Powered Motor And Accessory/Attachment
|
Back to Search Results |
|
| Catalog Number |
451570102 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/17/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during set up for a total knee arthroplasty (tka) surgery, the robotic-assisted solution saw handpiece was leaking oil.A different saw handpiece was used during the surgery.Patient treatments were not delayed or cancelled.There was no patient involvement.The device was operated in conjunction with the robotic-assisted solution base station device and the robotic-assisted solution satellite station device.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11: additional narrative:
the actual device has been returned and is currently pending evaluation.Once the device has been evaluated, a supplemental medwatch report will be sent accordingly.
|
| |
|