INTUITIVE SURGICAL, INC None; ENDOSCOPE PLUS Back to Search Results
Model Number 470057-11
Medical Device Problem Code Image Orientation Incorrect (1305)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/02/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a da vinci-assisted surgical procedure, the 30-degree endoscope plus image was being displayed as a double image and was inverted.The technical support engineer (tse) viewed the system logs and found related errors on the endoscope, then advised the customer to replace the endoscope.The customer confirmed that the replacement endoscope resolved the issue.The procedure was completed with no reported injury.
 
Additional Manufacturer Narrative
An investigation was completed to determine the cause of this reported event.The reported event was addressed with phone support.Technical support engineer (tse) advised replacing the 30-degree endoscope plus.Once the new scope was installed, the customer did not encounter any further issues and proceeded with the case as planned.The system was working properly, and no additional action was required as the issue was resolved.Intuitive surgical, inc.(isi) did not receive a da vinci product involved with this complaint to perform failure analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
None
Common Device Name
ENDOSCOPE PLUS
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 95051
+1(408)523-2100
MDR Report Key50901511
Report Number2955842-2026-39750
Device Sequence Number18831023
Product Code NAY
UDI-Device Identifier00886874116555
UDI-Public(01)00886874116555
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K191736
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 09/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number470057-11
Device Catalogue Number470057
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 09/02/2026
Initial Report FDA Received Date09/25/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/08/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
da Vinci Instruments and Accessories.
Patient SexUnknown
-
-