The customer reported that the item was taken out of the box and it was not working, it was not holding air, and the cells in the mattress aren¿t inflating and when they do they are not keeping the air.The sample was received; the pump that came with the mattress was attached and the mattress did not fully inflate and the low pressure light on the pump did not turn off, then a known good pump was attached and again the mattress would not fully inflate and the low pressure light would not turn off.The reported issue was confirmed, and no holes in the mattress or issues with the connections were found.The cause was documented as cannot determine cause with received sample; a replacement was sent to the customer and the division trends the complaints quarterly.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|