It was reported that the surgeon was inserting a prophylactic long tfna nail into a patient.During insertion of the nail it was noted that the patient had hard bone and the nail was difficult to insert.The issue occurred when they used the 6/9mm step drill bit ¿ 03.037.022.The tip of the drill bit broke off in the patient after the drill bit was passed through the insertion handle, aiming arm and nail.There was no issues with insertion of the 10mm cannulated drill (03.037.021) bit prior to the 6/9mm step drill bit.The staff switched out the insertion handle, aiming arm and other relevant instrumentation and the case was completed without incident.The fragments in the patient were able to be retrieved.The 3.2mm guide wire (357.399) was bent when they inspected it after removing it from the patient.
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative:
d4: udi: as the catalog/model number was not provided, the (01)gtin is not available.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|