The account reported that the pilot ballon and extended tubing fell off while the patient was intubated; the cuff did indeed deflate and the patient was successfully extubated without harm to the patient.Medline quality received one unit in used condition without the original packaging, found the markings at the "18" and "20" graduation marks appeared to have rubbed off and are barely visible, and found the inflation tubing was not connected to the main et tube; at the base of the tube there was a yellow discoloration but no tearing or signs of breakage, and the inflation tubing could be placed back into where it would attach to the main et tube and freely come out as if the tubing fell out but did not break off.The returned sample confirmed the reported detachment, and based on the sample and complaint details the root cause is considered to be a vendor product issue relating to the manufacturing process; the vendor performed testing of the inflation tubing and found no issue in retention stock but stated that the most likely root cause was insufficient bonding.This complaint has been provided to the manufacturing facility for further evaluation, and the quality team will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|