| Model Number |
DDPB3D1 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Misconnection (1399); Over-Sensing (1438); Pacing Problem (1439)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/03/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the right ventricular (rv) lead triggered lead integrity alert (lia) for sensing integrity counter (sic) and high rate non-sustained episodes and had exhibited noise and oversensing resulting in inhibition of pacing.It was also stated that the implantable cardioverter defibrillator (icd) exhibited connection issue.A chest xray indicated that the device and rv lead exhibited connection issue as the rv lead was not inserted past the setscrew.The lead was reposition in device connector port and the device and lead remains in use.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|