Medtronic Europe Sarl Cobalt¿ DR MRI SureScan¿; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT) Back to Search Results
Model Number DDPB3D1
Medical Device Problem Codes Signal Artifact/Noise (1036); Misconnection (1399); Over-Sensing (1438); Pacing Problem (1439)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/03/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the right ventricular (rv) lead triggered lead integrity alert (lia) for sensing integrity counter (sic) and high rate non-sustained episodes and had exhibited noise and oversensing resulting in inhibition of pacing.It was also stated that the implantable cardioverter defibrillator (icd) exhibited connection issue.A chest xray indicated that the device and rv lead exhibited connection issue as the rv lead was not inserted past the setscrew.The lead was reposition in device connector port and the device and lead remains in use.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
  medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Cobalt¿ DR MRI SureScan¿
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
Medtronic Europe Sarl
route du molliau 31
case postale
tolochenaz vaud 1131
SZ  1131
Manufacturer (Section G)
Medtronic Europe Sarl
route du molliau 31
case postale
tolochenaz vaud 1131
SZ   1131
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112--439
+1(763)526-9277
MDR Report Key50902720
Report Number9614453-2026-03879
Device Sequence Number10762247
Product Code LWS
UDI-Device Identifier00763000178376
UDI-Public(01)00763000178376(17)270228(20)02
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
P980016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberDDPB3D1
Device Catalogue NumberDDPB3D1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/04/2026
Initial Report FDA Received Date09/25/2026
Date Device Manufactured09/09/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
5076-52 LEAD, 694765 LEAD
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age38 YR
Patient SexMale
Patient Weight280 LB
-
-