MAUDE Adverse Event Report: Tandem Diabetes Care t:slim X2 Insulin Pump with Control-IQ+ Technology; Alternate Controller Enabled Insulin Infusion Pump

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Tandem Diabetes Care t:slim X2 Insulin Pump with Control-IQ+ Technology; Alternate Controller Enabled Insulin Infusion Pump Back to Search Results
Model Number 1000354
Medical Device Problem Code Pumping Stopped (1503)
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 09/22/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that pump displayed a malfunction alarm related to touchscreen and showed a malfunction code, and customer¿s blood glucose reading displayed ¿high¿ with a value above 500 mg/dl.Customer¿s blood glucose level exceeded 500 mg/dl, but specific value was not known.Customer gave correction insulin by injection and did not need help from emergency services, and technical support provided pump reset information, assisted customer with resetting malfunction alarm, confirmed alarm did not recur, advised customer to continue using pump, and instructed customer to call back if malfunction appeared again.
 
Additional Manufacturer Narrative
In use at the time of the event:\ \ cgm model: unknown.\ \ mobile os: unknown / unknown / unknown / unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
t:slim X2 Insulin Pump with Control-IQ+ Technology
Common Device Name
Alternate Controller Enabled Insulin Infusion Pump
Manufacturer (Section D)
Tandem Diabetes Care
12400 high bluff drive
san diego CA 92130
Manufacturer Contact
john connell
12400 high bluff drive
san diego, CA 92130
+1(858)224-6281
MDR Report Key50902829
Report Number3013756811-2026-175384
Device Sequence Number10886858
Product Code QFG
UDI-Device Identifier00389152407319
UDI-Public(01)00389152407319(22)1016562(11)250922
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K232380
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Type of Report Initial
Report Date (Section B) 09/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number1000354
Device Catalogue Number1014073
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/23/2026
Initial Report FDA Received Date09/25/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/22/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age18 YR
Patient SexUnknown
Patient Weight125 LB
-
-