| Model Number |
2120 |
| Medical Device Problem Code |
Device Sensing Problem (2917)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/06/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The suspect pump was returned for evaluation.The event history log (ehl) was reviewed.A review of the 12-month service history was conducted; however, no relevant findings were identified.Visual inspection and functional testing were performed.The customer-reported issue was confirmed through the downstream occlusion test.The cause of the reported issue was determined to be a defective dso sensor.Dso sensor was replaced to resolve issues.Performance verification testing (pvt) was performed, and the unit passed all testing criteria.
|
| |
|
Event or Problem Description
|
|
It was identified during the inspection of the returned pump that the event log contained the error message, ¿cassette not attached properly.¿ this high-priority error was observed during testing.Although no issues were reported during patient use, recurrence of this error during an infusion could interrupt therapy.There was no patient involvement associated with this event, and no patient harm was reported.
|
| |
|